知識社群
登入
中文(台灣)
English(US)
Q&A
線上人數:
427
MODERN RADIOTHERAPY AND THERAPEUTIC RADIOLOGICAL PHYSICS
相簿輪播 - Welcome Address
社群功能
Home page
Documents
(15)
Calendar
Photo Album
(4)
文件分類
[
總覽
]
Sri Lanka WS
(15)
Vietnam WS
社群資訊
訪客: 71013
文章: 15
討論: 0
公告: 2
容量: 剩餘
1.4 GB
(1.9 GB)
閱讀權限: 開放
分類:
醫療保健 / 醫學新知
版主: tcc
副版主: 無
位置:
MODERN RADIOTHERAPY AND THERAPEUTIC RADIOLOGICAL PHYSICS
>
Documents
>
Sri Lanka WS
Dosimetry In-vivo Dosimetry
(33:26)
Tweet
by tcc, 2011-04-21 18:51, 人氣(5125)
嵌入
|
cc Media
|
轉檔資訊
更清晰的畫質或聲音,請點選(限 IE)
標題
1.
In-vivo dosimetryKrishna Komanduri PhDDirector of PhysicsCancer Treatment Services International
2.
Definition of the Problem
3.
Slide 3
4.
Slide 4
5.
For High-energy electrons Surface dose Target dose Total skin irradiation
6.
Slide 6
7.
Slide 7
8.
Slide 8
9.
Slide 9
10.
Slide 10
11.
Slide 11
12.
Slide 12
13.
Slide 13
14.
Slide 14
15.
Practical consequencesThermal fading should be evaluated on each individual reader with the TLmaterial which is intended to be used. It should be approximately 1% per month, orless, for the different preparations of LiF when readout and annealing conditions arereached. For Li2B4O7 it varies from 0,5 to 1% per week depending upon the doping.When a long delay separates irradiation from readout, a fading correction maybe necessary.TL materials which show a standard deviation higher than + 2 % are either ofpoor quality or are not correctly handled and read out. They should not be used for invivo measurements or the handling procedure should be improved.
16.
Slide 16
17.
Practical consequencesThermal fading should be evaluated on each individual reader with the TLmaterial which is intended to be used. It should be approximately 1% per month, orless, for the different preparations of LiF when readout and annealing conditions arereached. For Li2B4O7 it varies from 0,5 to 1% per week depending upon the doping.When a long delay separates irradiation from readout, a fading correction maybe necessary.TL materials which show a standard deviation higher than + 2 % are either ofpoor quality or are not correctly handled and read out. They should not be used for invivo measurements or the handling procedure should be improved.
18.
Slide 16
19.
Slide 17
20.
Slide 18
21.
Practical consequencesIn practice it is recommended to use TL dosimeters in the region where theirresponse is proportional to the dose received (linear region). When it is not the case, a correction should be applied to the signal from a curve established with the TL material as well as the reader used (and not from a published curve because the readout parameters may have an influence on its shape). This curve should be checked periodically.It can be introduced in the computer associated with some automatic readout systems, thus allowing an automatic correction of the readouts. TL dosimeters should not be used in the sub linear region approaching saturation.
22.
Slide 20
23.
Slide 21
24.
Slide 22
25.
Slide 23
26.
Slide 24
27.
Geometrical Configurations:
28.
Slide 26
29.
Diodes …contd Influence of temperatureInfluence of energy Photon energy Electron energy Directional effect
30.
Clinical use Calibration factors of a diode Entrance dose calibration factor Exit dose calibration factor Use of a diode in clinical conditions
31.
Slide 29
32.
Slide 30
33.
Slide 31
34.
(1). Entrance Calibration Factor, Fcal,en
35.
(2). Exit Calibration Factor, Fcal,ex
36.
Beam dependent correction factors
37.
Intrinsic correction factors
38.
Optional study:
39.
Slide 37
40.
Practical consequences:Diodes and associated electrometer should have a drift less than 1% in the time range encountered in practice. Equipment falling outside this range should notbe accepted.
41.
Slide 39
42.
MOSFET dosimeters
43.
Slide 41
44.
Slide 42
45.
Slide 43
46.
Slide 44
47.
Slide 45
48.
Slide 46
49.
Slide 47
50.
Slide 48
51.
Slide 49
52.
Slide 50
53.
Systematic Problem with QA
54.
Workflow -- Visualize, Verify, Validate
55.
... [more]
© 2010 FormosaSoft Corp.
http://www.powercam.com.tw
service@powercam.com.tw
All rights reserved.